User Guide
17
RECOMMENDED STIMULATION SCHEDULE
Use the TrueTear
®
device at least twice a day, as needed. Stimulation longer than
3 minutes is not recommended, and you should wait for at least 60 minutes before
proceeding to the next application. The device has a built-in single-day usage limit
of 30 minutes. If this daily limit has been reached, the TrueTear
®
device will turn on
and then off immediately. The device will not deliver stimulation.
Replace the tip every 28 days with a new tip. When a tip has 7 days of usage left before
expiration, the device will vibrate 3 times when the device is turned on or off. If the device
is turned on without a tip or with an expired tip, the device will vibrate 3 times but will not
deliver stimulation.
WHAT YOU'LL SEE WHAT IT MEANS
A single, steady white light when device
is turned on.
Device is on. No stimulation is delivered.
A single, ashing white light when the
device is turned on.
Battery is running low. Place base in case.
All lights ash on and off. Daily stimulation limit of 30 minutes has
been reached.
Figure 9. Cleaning with alcohol wipes.
CARING FOR YOUR TRUETEAR
®
DEVICE
1 Use alcohol wipes to clean the device and store
the device in the case between uses.
2 Use alcohol wipes to clean tthe case as needed.