User's Manual
User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 81
variations on
power supply
input lines IEC
61000-4-11
cycles
Voltage
interruptions:
0 % U
T
; 250/300
cycle
Voltage
interruptions:
0 % U
T
; 250 cycle
If the user of the
T4336RWC and
T4336RWG requires
continued operation
during power mains
interruptions, it is
recommended that the
T4336RWC and
T4336RWG be powered
from an uninterruptible
power supply or a
battery.
Power
frequency(50, 60
Hz) magnetic
field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz
The T4336RWC and
T4336RWG power
frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
professional healthcare
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Manufacturer’s declaration-electromagnetic immunity
The T4336RWC, T4336RWG is intended for use in the electromagnetic environment (for
professional healthcare) specified below.
The customer or the user of the T4336RWC, T4336RWG should assure that it is used in
such and environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment
-guidance (for professional
healthcare environment)
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80
3 Vrms:
0,15 MHz – 80
Portable and mobile RF
communications equipment should