User's Manual
User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 80
Manufacturer’s declaration-electromagnetic immunity
The T4336RWC, T4336RWG is intended for use in the electromagnetic environment (for
professional healthcare) specified below.
The customer or the user of the T4336RWC, T4336RWG should assure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-guidance
(for professional
healthcare environment)
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power
supply lines
+ 1kV for
input/output lines
+ 2kV for power
supply lines
+ 1kV for
input/output lines
Mains power quality
should be that of a
typical professional
healthcare environment.
Surge IEC
61000-4-5
+ 0.5kV, +1kV
line(s) to line(s)
+ 0.5kV, +1kV,+ 2kV
line(s) to earth
+ 0.5kV, +1kV
line(s) to line(s)
+ 0.5kV, +1kV,+ 2kV
line(s) to earth
Mains power quality
should be that of a
typical professional
healthcare environment.
Voltage Dips,
short
interruptions
and voltage
Voltage dips:
0 % U
T
; 0,5 cycle
0 % U
T
; 1 cycle
70 % U
T
; 25/30
Voltage dips:
0 % U
T
; 0,5 cycle
0 % U
T
; 1 cycle
70 % U
T
; 25 cycles
Mains power quality
should be that of a
typical professional
healthcare environment.
Voltage fluctuations
/flicker emissions IEC
61000-3-3
Compliance
low-voltage power supply network that
supplies buildings used for domestic purposes.