User's Manual
User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 79
4. The use of accessories, transducers and cables other than those specified,
with the exception of transducers and cables sold by Allengers Medical
Systems Ltd. as replacement parts for internal components, may result in
increased emissions or decreased immunity of the detector.
5. The detector should not be used adjacent to or stacked with other
equipment.
If adjacent or stacked use is necessary, the Allengers FPD should be observed
to verify normal operation in the configuration in which it will be used.
6. Basic performance of the equipment and the system
After image data are acquired from the flat panel detector, offset data
correction is performed by flat panel detector and then the image is send to
DR system via wireless connection or wired connection.
7. Test items
Manufacturer’s declaration-electromagnetic emissions
The T4336RWC, T4336RWG is intended for use in the electromagnetic environment (for
professional healthcare) specified below.
The customer or the user of the T4336RWC, T4336RWG should assure that it is used in
such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
(for professional healthcare environment)
RF emissions CISPR 11
Group 1
The T4336RWC, T4336RWG uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class A
The T4336RWC, T4336RWG is suitable for use
in all establishments other than domestic and
those directly connected to the public
Harmonic emissions
IEC 61000-3-2
Not
applicable