- Connect the equipment into an outlet on a circuit different from that to
which the other device(s) are connected.
If the problem cannot be solved with the above measures, stop using this equipment
and consult the manufacturer, our Allengers dealer or Representative for help.
Warning
Do not place devices generating electromagnetic wave near this equipment.
If a device(s) other than those specified is connected, predetermined EMC
performance cannot be guaranteed.
B.2 Further Information for IEC 60601-1-2 (EN 60601-1-2)
Caution
This equipment is not intended for use in residential environments and may
not provide adequate protection to radio reception in such environments.
1. Medical electrical equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information
provided in the accompanying documents.
2. Portable and mobile RF communications equipment can affect medical
electrical equipment.
3. Information regarding the cable affecting EMC is as follows.
Name
Connection
Maximum
length
General
specification
Network
cable
Extend
Back-up
cable to PC
or network
switch.
20m
Cat 5e or more
UTP type and
straight cable
Power
cord
From
adaptor to
AC outlet
Use a hospital grade power cord