User's Manual
User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 75
Item
Min
Typ.
Max
unit
MTF @ 1 lp / mm
0.50
0.63
DQE @ 1 lp / mm
0.41
0.55
T4336RWG (GOS Version) Performance at *RQA5
Item
Min
Typ.
Max
unit
MTF @ 1 lp / mm
0.46
0.52
0.57
DQE @ 1 lp / mm
0.23
0.27
0.32
Sensitivity
503
592
681
LSB/uGy
Dark noise
4.8
LSB
Maximum linear
dose
75
88
101
uGy
Energy range
40
150
KVp
Maximum
resolution
3.57
lp/mm
T4336RWG (GOS Version) Life Time Performance at *RQA5
Item
Min
Typ.
Max
unit
MTF @ 1 lp / mm
0.44
0.52
DQE @ 1 lp / mm
0.22
0.27
*The above specifications have been verified by Allengers; The physical
characterization of the systems was obtained with the standard beam conditions
RQA5 (in accordance with IEC 61267:1994).
A.1.7 Radio Waves
Wireless specifications for the flat panel detector and the access point are as follows.