User's Manual
User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 73
Functional means that the detector is able to deliver an image, whereas the image
quality and electro-optical performances are not guaranteed. It also means that,
except if mentioned explicitly, any perturbation occurring within those conditions
are reversible and not existing anymore when the detector is turned back within the
performance operating conditions.
Item
Min
Typ.
Max
Temperature
5°C
25°C
35°C
Humidity
15%
80%
Pressure
700mBar
1013mBar
1060mBar
Weight (uniform load)
300 kg
Weight (local load on 4 cm
diameter)
120 kg
Vibrations (10 -200 Hz / 3
axis )
2 g
Drop height (the
reversibility to full
performance is not
guaranteed for this item)
70 cm
A.1.5.3 Performance Operating
Performance is guaranteed in the operating range below.
Item
Min.
Typ.
Max.
Temperature
15°C
25°C
30°C
Humidity
15%
80%
Pressure
700mBar
1013mBar
1060mBar
Distance to calibration
temperature
-12°C
0°C
+12°C
Weight (uniform load)
300 kg
Weight (local load on 4 cm
diameter disc)
120 kg
Vibrations (10 -200 Hz / 3
2 g