User's Manual

User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 55
Image module: Alarm caused by image module
Storage: Alarm caused by storage failure
Ethernet: Alarm caused by Ethernet interface failure
Wireless: Alarm caused by Wireless interface failure
Self-diagnosis: Alarm caused by software self-diagnosis failure
Control Host: Host name of the PC currently controls the operation of FPD
Error (The group is collapsed by default, click + to expand)
Error Code: A 32-bit mask of following error flags (0x00000000
means no error)
ROIC 0~17: Failure on checking ROIC index N (0~17) configuration
(bit 0 ~17)
Gate: Failure on checking gate (bit 24)
ROIC init: ROIC initialization timed out (bit 25)
AED: Failure caused by AED operation (bit 26)
DMA: DMA timed out (bit 27)
Power: Power system failure (bit 28)
Image module: Image module timed out (bit 29)
Panel: Panel configuration error (bit 30)
Driver: Software driver error (bit 31)
Configurations
Select items from dropdown menu
Update: Retrieve corresponding information from FPD and update the
displayed information
Set: Save the configuration to FPD. (Reboot is required for applying
configurations to FPD)
Ethernet (Reboot is required after Set action)
IPv4: IPv4 of FPD
IPv4 prefix: 24 (fixed)
IPv4 gateway: Gateway setting
IPv4 DHCP: Disable / Client / Server
IPv6 functions not yet implemented