Set FPD time: Synchronize the time between PC and RTC of FPD.
Mode
Start: Make FPD operate in selected exposure mode.
Stop: Make FPD stop selected exposure mode and return to Idle in operation state.
Reset: Reset the image module (Clear some alert flags).
Mode: Choose exposure mode (AED / AED repeat / Soft Sync / Acquisition /
Discharge).
AED: Automatic exposure detection is activated with Utility, and to detect the X-ray
and capture image directly.
AED repeat: Automatic exposure detection is activated with Utility at first time, the
following activation will be done with panel itself and saved in the storage of FPD. *
Soft Sync: Synchronizing detector and radiation system to exposure.
Acquisition: Capturing image without X-ray radiation.
Discharge: Reset the panel.
* Alarm: Please unchecking “don’t save to eMMC” while using AED repeat mode, or
the images would not be saved properly once the temporary storage of DDR is full.
User Key: A number for user key generation. The number increases by 1 after each
exposure.
Download Raw: Download the raw image to the target folder.
Download Offset: Download the offset image to the target folder.
Download Corrected Image: Download the corrected image to the folder named
“Corrected”. (It will automatically create the folder named corrected but the
calibration is depending on the availability of the maps or offset configuration.)
Preview Image: Activate the function of preview image and download the preview
image to the folder named “ThumbNail”. (The saving priority is the first if checked)
Don’t Save to eMMC:
When this option is not checked, the image is saved in the non-volatile storage of
FPD.
When this option is checked, the image is only stored in buffer of FPD and will be
overwritten by subsequent exposure image.
Acq Repeat: The numbers of images would automatically be collected. (Only for
images without exposure.)
Bulk Size: Determine the data size to be divided to transfer.
Update Maps: Renew the maps in the Calibration Map folder to process the
calibration
Sub-folder: Provide the options to download the images in the corresponding
subfolder.