5.2 Detector Connection
Allengers detector can be connected to system either by wireless or by wire
cable. Allengers FPD use Dynamic Host Configuration Protocol (DHCP) to assign
IP address automatically.
5.2.1. Wire connection
The default address of detector is 192.168.11.100. User need to set the
computer to the same domain as Allengers. First, open the IPV4 of
computer's local network management interface, and change the IP address
to 192.168.11.1 and subnet mask to 255.255.255.0.
After connection established, open Cmd.exe and enter ‘ping 192.168.11.100’
then press enter to check the network. If success, there is a reply from FPD