User's Manual

User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 28
a patient.
The institution where the equipment is installed is responsible for its use and
maintenance. In addition, this equipment should not be used by persons
other than doctors or suitably trained staff.
Be careful not to expose the flat panel detector to X-ray without a subject.
Although the flat panel detector conforms to IPX6, no warranty is given as to
the prevention of water intrusion in the flat panel detector. If the flat panel
detector is splashed with water, wipe off moisture and ensure that the flat
panel detector is completely dry before use.
As the cables of the equipment are long, be careful not to entangle the cables
during use. Also, be careful not to trip over the cables. Falls could result in
injury.
Follow the specified procedure when turning off the equipment. Otherwise,
the flat panel detector could be damaged by thermal shock.
External equipment intended for connection to signal inputs, signal outputs
or other connectors shall comply with the relevant product standard e.g. IEC
60950-1 or IEC 62368-1 for IT-equipment and the IEC 60601-series for
Medical Electrical Equipment. In addition, all such combinations Medical
Electrical Systems shall comply with the safety requirements stated in the
general standard IEC 60601-1, edition 3, clause 16. Any equipment not
complying with the leakage current requirements in IEC 60601-1 shall be kept
outside the patient environment i.e. at least 1.5 m from the patient support
or shall be supplied via a separation transformer to reduce the leakage
currents.
Any person who connects external equipment to signal inputs, signal outputs
or other connectors has formed a Medical Electrical System and is therefore
responsible for the system to comply with the requirements. If in doubt,
contact qualified medical technician or your local representative.
Do not place FPD near strong EMI sources, it might cause line defect in the