the equipment or battery charger, and stop using it. Otherwise, the battery
pack may overheat, emit smoke, exploder ignite.
Do not use the battery pack exposed to a strong magnetic field of an MRI
system, etc.
Do not use the battery pack immersed in liquid.
Keep the battery away from fire. Avoid the action of short circuit. Never
attempt to disassemble the battery pack.
Use the designated chargers to charge the battery pack. Risk of explosion if
battery replaced by an incorrect type.
Dispose of used batteries according to the instructions.
When using the FPD for the first time, because the internal battery may be
empty, do not install the main battery, but use the wire to charge the FPD
internal battery, and charge it for at least 15 minutes before it can be used
normally.
Regarding the storage of the battery, first of all, the best storage capacity of
the battery is about 40~50%. Before the installation, when the battery and
the product are still in the warehouse, it is recommended that the main
battery and the internal battery must be checked every 3 months to charge.
The main battery needs to be charged to 50%, two bars of the charging stand,
and it takes about 2 hours to charge from 0%. ; The internal battery is also
charged to 50%, you can use the UI to check the level of the internal battery,
it takes about 1 hour to charge from 0%.
After charging, make sure that the hotswap function operates normally,
please unplug the main battery and press the power button to test the
internal battery power at least 3 lights, and the internal battery power should
be more than 50%