User's Manual
User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 21
Do not use the following disinfectants at the time of disinfection, which the
quality, performance and safety of the equipment cannot be assured.
- Chloric disinfectant which is strongly corrosive to metals and rubber parts.
- Disinfectant whose uses on metals, plastics, and coating are forbidden
according to the instructions supplied with the disinfectant.
- Formalin gas and disinfectant sprays that may get inside the equipment.
- Ultraviolet sterilizers.
Disinfectant ethanol is recommended for disinfection. Carefully read the
instructions and cautions supplied with the disinfectant before use.
For details on the disinfectant, contact a Allengers dealer or the service
representatives at the agency from which you purchased the disinfectant.
Caution
If flat panel detector is not disinfected, it may lead secondary infection.
Be sure to disinfect with ethanol after use.
Clean the flat panel detector with ethanol for disinfection, etc. for each
patient to prevent infection.
3.11 Charging the Battery Pack
Caution
Charging battery only by the back-up cable provided by Allengers when
battery is installed on Allengers flat panel detector or use the battery charger
provided by Allengers medical Systems Ltd. For details on operations, refer to
the instruction manual for the battery charger.
Do not charge the battery pack near fire or under strong sunshine. If the
built-in protection mechanisms are activated by a high temperature, the
battery pack cannot be charged. Also, if the built-in protection mechanisms
are damaged, the battery pack may be charged with extremely high current
and voltage, and abnormal chemical reactions may occur inside the battery