For veterinary or mobile applications, contact Allengers official dealer or
Representative.
When the devices are used outdoors in wireless communication mode,
contact Allengers official dealer or Representative.
Do not place any object in a place where removal of the power cable is
prevented.
To ensure optimal image quality, it is recommended that you do not use the
flat panel detector near devices (motor, transformer, switching supply, etc.)
that generate electromagnetic noise.
To ensure optimal image quality, it is recommended that you do not place
the cables (power cable, back-up cable, etc.) of the equipment near devices
(motor, transformer, switching supply, etc.) that generate electromagnetic
noise and their cables.
Please use the engineering version of UI to reset the Real-time clock (RTC)
every time you use it, or make sure that the RTC of the FPD is correct when
connecting to the radiology software, to avoid wrong time records in the log.
If you use the engineering version of the UI for single-dose calibration, it is
recommended to use 70kv and the raw image to reach the conditions of
30000±3000lsb to achieve the best calibration map.
Please use the Allengers service tool to update the FPD firmware to avoid
update errors.
3.6 Connection Instructions
Warning
Make sure that the devices to be connected to the equipment are authorized
for connection.
Connect the panel unit Flat Panel detector only to the access point or DR
system.