User's Manual
User Manual
Manual No. AMSL/DMR/UM/FD 009
Rev. 00 Effective Date 12.07.2022 Page | 10
2.4 Product Significant Performance Characteristics
Items
Characteristics
Image resolution
2500x3052
Data
16 bit data
MTF (Modulation transfer
function) (Typical)
CsI
0.63 @ 1 lp/mm
GOS
0.52 @ 1 lp/mm
DQE (Detector quantum
efficiency) (Typical)
CsI
0.50 @ 1 lp/mm
GOS
0.27 @ 1 lp/mm
Maximum linear dose (Typical)
88 uGy
Sensitivity (Typical)
CsI
574 lsb/uGy
GOS
592 lsb/uGy
Dark noise (Max)
CsI
3.6 LSB
GOS
4.8 LSB
Lag
<0.5% after 1 minute
*Remark: There are 200 bytes come with each image, including 96
bytes for header and 104 bytes for footer.
2.5 Product Intended Patient Population
Considerations
Requirement Description
Age
- Not restricted
Height
- Not restricted
Weight
- The uniform load of the device is
300kg
- The local load on 4 cm diameter of
the device is 120kg.
*Note: Please evaluate the patient’s
condition before using.
Health
- Not restricted
Nationality
- Multiple
Sex
- Not restricted