Use Instructions
Instructions for Use for KardiaMobile® (Model AC-019)
Page 17 of 23
○ Option 5: If you wear hearing aids, turn them off prior to recording.
○ Option 6: Avoid close proximity to items that may cause electrical interference
(electronic equipment, computers, chargers, routers, etc.)
3. Problem: The ECG rhythm from Lead II does not appear while recording.
Solution
○ Ensure that the bottom electrode (Left Leg Electrode) is touching the skin above your
left knee or your left ankle.
NOTE: It is recommended that the bottom electrode (Left Leg Electrode) touches
bare skin when you record a Six-Lead ECG.
4. Problem: The ECG rhythms appear upside down.
Solution
○ Option 1: Make sure the AliveCor logo is in the correct orientation. Make sure your
thumbs are touching the 2 top electrodes (left thumb on the Left Hand Electrode and
right thumb on the Right Hand Electrode) and the bottom electrode (Left Leg
Electrode) is touching the skin above your left knee or your left ankle.
NOTE: It is recommended that the bottom electrode (Left Leg Electrode) touches
bare skin when you record a Six-Lead ECG.
○ Option 2: Invert the ECG.
In the Instant Analysis screen, tap “Invert” on the top left corner to toggle it ON or
OFF.
In the ECG Review screen, tap “MORE” at the bottom of the ECG Review
screen, and then tap the “INVERT” switch to toggle it ON or OFF.