Owner manual


Specication
Meter or
Gauge
Small Valve
Controller
Large Valve
Controller
Description
Input /Output Signal Digital PROFIBUS DP
Electrical Connections DB9
Supply Voltage: 7 to 30 Vdc 12 to 30 Vdc 24 to 30 Vdc
Supply Current
80mA @ 12Vdc
65mA @ 24Vdc
295mA @ 12Vdc
280mA @ 24Vdc
780mA @ 24Vdc
Technical Data for PROFIBUS Meters, Gauges and Controllers
NOTICE: The following specications are applicable to Alicat PROFIBUS enabled meters,
gauges and controllers only.
All other operating specications are shown in the Technical Data page for standard Alicat
instruments.
All standard device features and functions are available and operate in accordance with the
standard Alicat Scientic device operating manual provided with the device.
Power and Signal Connections:
Connect to the device using two DB9
connectors.
The female top connection is PROFIBUS.
The male connection on the side is power and
RS-232 or RS-485.
Pin out diagrams for all PROFIBUS enabled
Alicat devices are shown:
PROFIBUS MC1SLPM shown to provide PROFIBUS connector dimensions only. Flow
body and valve dimensions will vary with range. Please see Alicat’s device specications for
complete dimensions.
PROFIBUS units do not have a display screen.
2 x 8-32 UNC
v
.350
.525
.35
4.26
4.08
3.76
3.59
1/8 NPT
Both Sides
1.05
.525
.35
2.225
.150
.125
.925
MC-1SLPM-PROFIBUS
1. NC
2. OPT GND
3. DP
4. RTS
5. DGD
6. VP
7. OPT 7 to 30VDC
8. DN
9. NC
5 1
9 6
TOP
1. NC
2. RS232RX
3. RS232TX
4. NC
5. GND
6. NC
7. 7 to 30VDC
8. GND
9. NC
1
6
5
9
SIDE