Specifications

56
Network Communications—N2 Communications Bus
0 V -
Ch 1 = 2 V/div.
T/div 0.5 mS
3 to 10 V pp
(including +0.5 VDC Bias)
scopesg
Figure 33: N2Bus Signal as Measured with Oscilloscope
6.
Connect Channel B to the N2- wire and connect Channel A to the
N2+ wire. Connect the oscilloscope ground lead to the N2 Reference
wire.
7.
Compare the signal you measure with the waveform in Figure 33. They
should resemble the figure. A signal outside this range indicates that
the N2 Bus is not functioning properly.
Note: At the dead time between messages, you can measure the bias
voltage. In this figure, it is 0.5 VDC.
Factors that can cause improper signal levels are:
N2 Bus wires are not polarized correctly (i.e., wires are reversed).
N2 cable runs are too long.
24 AWG wire has breaks in it.
An ASC has malfunctioned.
N2 Bus is improperly terminated (e.g., EOL jumpers configured
improperly).
N2 Bus contains too many devices, loading down the bus.
N2 Bus wiring has T or Y connections installed.
Check all wiring and devices to pinpoint the problem. Replace parts if
necessary.