9.2.3 Fixation Test
The fixation test evaluates the previous target selection if the user is capable and or willing to look at
the target. This tests confirms that there is some kind of reaction of the users to the target. It can reveal
brain neglects as well as visual impairments.
The test will switch the gaze target between two spots on the monitor and repeat this three times. It
will measure if the eyes meaningfully respond to this stimulus change.
Fixation Test Parameter Description
Precision OUTPUT:
Left / Right
A low value < 100 px indicates that the user did
react to the left or right stimulus.
Average fixation OUTPUT: ms How long was the user able to fixate the targets.
This value a also an early indicator how long the
maximum dwell time should be
Perform Fixation
Test
Starts the test.
Gaze-Ass. Hotkey Function Comment
Enter
Starts the test
Escape
Interrupts the test
78
Version: 3.0
© alea technologies gmbh, 2015
Bad – arbitrary reaction on the left spot
Good – clear reaction on both spots