Installation Instructions

IntelliGaze™ Installation & Service Manual IntelliGaze™ Reference
Software Version & Online Update
This entry shows the currently installed software version and can optionally provide information about
available online updates.
Please refer to chapter 8 'IntelliGaze Software Update' for details.
6.5.2 Calibration
During the calibration, the IntelliGaze™ system calibrates the internal calculations to the individual
parameters of a subjects eye.
A calibration requires the subject to fixate a number of targets on the screen. Typically the
system will automatically detect the correct fixation and interactively cycle through the
point sequence. The parameters for this process can be configured as described below.
After the calibration, the system will provide an intuitive feed-back about the quality of the calibration.
Once a calibration has been completed successfully, it will be automatically saved and re-used, until
another calibration has been completed. This works even over restarts of the system.
In order to achieve and maintain the highest accuracy, we recommend calibrating the user after system
changes or longer periods of time. Very large changes in ambient illumination might also be compensated by
a new calibration run.
The parameters of the calibration can be configured under: Options Calibration
© alea technologies gmbh, 2015
Version: 3.0
61