Installation Instructions

Setup IntelliGaze™ Installation & Service Manual
4.7 Monitor Calibration
The last step of the setup process is the calibration of the monitor-camera configuration
The monitor calibration requires a metric ruler of about 40 cm.
For further details please refer to chapter 5.4.
A badly calibrated monitor will result in inaccurate head-motion compensation and therefore provide inaccu-
rate gaze position measurements.
4.8 System Validation
This step will make sure that all components have been installed correctly and the system is working
according to specification.
It assumes all setup-steps above have been completed successfully.
1. Connect all cables.
At the back of the CAM30 system there is a LED that turns green if the illumination has power.
The LED of the CAM30NT shows the green LED when IntelliGaze is started.
2. Turn-on the PC and boot Windows. Login.
3. Start IntelliGaze™, if not included in 'Autostart'
You should see the following:
Dark red shining outer LEDs, the illumination will go to a standby mode if no head
was detected for a while. In standby mode, the outer LEDs will flash only about once
a second.
The status windows showing a green or yellow head symbol when the head is within
the working range
Ctrl + F5 start a calibration (if the monitor is not calibrated you will get a message
that the calibration can't be started)
4. If the system is calibrated you should see the cursor following your gaze.
30
Version: 3.0