User's Manual Part 5
6.2
8065751131
SLIT LAMPS WITH DOCTOR PROTECTION FILTERS AND ADAPTATIONS
A
slit lamp assembly is typically used to deliver the
PurePoint
™
laser treatment beam to the
™
laser treatment beam to the
™
patient's eye. An
adaptation, mounted on the slit lamp, is required to interface the slit lamp to
the
PurePoint™
laser.
PurePoint™ laser.PurePoint™
The adaptation consists of a zoom assembly with micromanipulator and a
Doctor Protection
Filter. The Doctor Protection Filter provides eye protection for the physician. The parfocal
zoom assembly is used to set the spot size of the aiming and treatment beams, and a
micromanipulator is provided for fi ne adjustment of the beam position.
Figure 6-1 Alcon SL1000 Slit Lamp
Spot size
adjusting knob
Spot size
adjusting knob
Figure 6-2
Zeiss 30SL