User's Manual Part 5

6.18
8065751131
12. Press the Standby/Ready key on the front panel. The green Standby LED turns OFF,
and the red Ready LED illuminates.
NOTE: The footswitch must be released to proceed to Ready mode. If the footswitch
is depressed during power-up or while in Standby mode, “Release footswitch”
is displayed. Release footswitch and proceed.
13. Use the Laser Vertical Adjustment Knob (see Figure 6-12) on the laser delivery
adaptation to aim the laser at the desired location within the illumination fi eld.
14. Press the footswitch when ready to fi re. The system will emit a 4 millisecond beep each
time the laser fi res. If the footswitch is not pressed within 2 or 10 minutes starting from
entry into “Ready” mode, the system emits one beep and switches to “Standby” mode.
NOTE: The aiming beam is off during treatment beam exposure, except in repeat mode.
15. Repeat the fi ring procedure as often as necessary, making adjustments to power output
and duration as appropriate to complete the treatment session.
16. When the treatment is completed, release the footswitch and press the Standby/Ready
key. The green Standby LED illuminates and the system is placed in “Standby” mode.
NOTE: You can disable both treatment and aiming lasers by pressing the Laser ON/
OFF switch. When turning the switch ON again, the system will default to the
last terminal selection used before shutdown with the exception that LIO will
default to Endo. Parameters shall be restored to the selected terminal.
Turn Off Sequence
1. Turn the Power Adjust knob to the minimum position.
2. Turn the key to the OFF (O) position and, for safety reasons, remove the key.
NOTE: The emergency switch on the front panel must only be used in case of
emergency. After using the emergency switch, pull it back to its initial position
to restore power and start the instrument.
3. Place the power switch on the rear of the system in the OFF (O) position.
NOTE: Between patients you can use the LASER ON/OFF switch to disable the
treatment and aiming beams. The cooling system remains active in this mode.
4. Place the illumination power switch to the OFF (O) position.