User's Manual Part 5

6.16
8065751131
Using the Alcon LIO-AT for Observation
If the LIO-AT is used for illumination purposes only, the laser fi ber does not need to
be connected to the
PurePoint
Laser
.
Note: Put dust cover on fi ber termination to
protect fi ber when not connected to
PurePoint
PurePoint
PurePoint
Laser
.
Laser.Laser
1. Turn the illumination power supply on.
2. Adjust the light intensity with the power supply illumination control knob.
Using the Alcon LIO-AT for Laser Treatment
Using the system in this mode enables photocoagulation with the LIO-AT.
WARNING!
All the personnel in the room during the operation must wear protective safety
eyewear with a minimum optical density OD 4 to fi lter 532nm radiation.
Before each use of the headset, the operator must examine the permanent Doctor
Protection Filter for scratches, breaks, or alterations by looking through the
ocular lens. If there is any doubt, discontinue use of device and please call Alcon
Technical Service.
NOTE: The LIO-AT is shipped with +2 diopter ocular lenses installed. These may be
changed with 0 (zero) diopter lenses.
1. If desired, change the ocular lenses by unscrewing the eyecup retainer in the
counterclockwise direction, change each lens, and replace the eyecup retainers. Ensure
that the new lenses are clean, i.e. no fi ngerprints or debris. Refer to the LIO-AT
maintenance section for cleaning instructions.
Figure 6-13
Eyecup Retainers and Ocular
Lens on the Alcon LIO-AT
2. Turn the
PurePoint
console power ON and make the appropriate selections as
specifi ed in Operator's Manual.
3. Turn the LIO-AT Illumination power supply ON.
4. Select the appropriate illumination fi eld size by toggling the illumination aperture lever
to the desired setting.
5. Adjust the illumination intensity using the power supply illumination control knob.