User's Manual Part 5

8065751131 6.15
8. Remove the LIO-AT and look at the scale on the eyepieces to insure that the pupil
distance is symmetrical. If not, center the headset and readjust the eyepieces. Correct
adjustment of the optics is particularly important when examining small pupils.
Once set, changing the adjustments is required only if another examiner uses the instrument.
Controls for Observation and Illumination
The Aperture Lever (see Figure 6-12) allows you to choose between two different-sized
illumination fi elds. The choice of illumination fi eld size depends mainly on the size of the
patient’s pupil (the small illumination fi eld is the recommended setting). The positions of the
Aperture Lever for large and small illumination fi elds are marked with large and small black
dots, respectively.
The Convergence Control Knob provides synchronized adjustment of both examination and
illumination beams to suit the patient's pupil size. Wide convergence and parallax selection
allows for maximum stereopsis with large pupils. Narrow convergence and parallax selection
allows stereoscopic examination for small pupils.
NOTE: Use the small pupil setting and
narrowest convergence angle at the small illumination fi eld size setting; otherwise,
clipping (shadow) of the illumination fi eld will occur. The Convergence Control Knob
adjustment range is limited in the LIO-AT to 50% of the original
Heine
range to
accommodate for the laser beam delivery requirements.
The Delivery Mirror Control Knob can be rotated to move both the illumination beam and the
laser beam in the vertical plane.
CAUTION
Do not use the LIO-AT with the illumination power supply set at maximum intensity
for
more than 10 continuous minutes. The LIO-AT must be allowed to cool down at
least 20
minutes between uses. Use as little observation/illumination light as possible and
always switch power supply OFF after use.
Circumference
Adjustment Knob
(on backside of band)
Observation Optics
Adjustment Knob
Laser Vertical
Adjustment Knob
Delivery Mirror
Control Knob
Height Adjustment Knob
Aperture Lever
Laser Spot Size Lever
Convergence Control Knob
Figure 6-12 LIO-AT Controls and Adjustments