User's Manual Part 5

6.14
8065751131
Using the Optics Overband
The pivoting overband allows the laser optics to be pushed up out of the operator's fi eld of
view (see Figure 6-11). It is locked in the end position and can only be released by pressing
the Overband Adjustment Knob.
To pivot the overband, press the Overband Adjustment Knob with the right hand and pivot
the overband into the desired position (up for the “rest” position and down for the “working”
position). When the unit is properly adjusted, the overband can be lowered into the same
pre-selected working position. Once set, changing the adjustments is required only if another
examiner uses the instrument.
Figure 6-11 Adjusting the LIO-AT Overband
Observation Optics Adjustment
1. Loosen the Observation Optics Adjustment Knob (see Figure 6-36) so that the
observation optics are free to move. The Observation Optics Adjustment Knob can be
unscrewed and reversed to the other side for left-handed operators. Remove dust cover
protecting delivery window.
2. Place the LIO-AT on your head and adjust the circumference and height using the
Circumference and Height Adjustment Knobs so that the headband is fi rmly positioned
but comfortable.
3. For convenience, use clothing clip to attach the fi ber/cable assembly to clothing.
4. Move the eyepieces as close as possible to your eyes and look at the light spot at a
distance of 30 cm. A small object (such as a pencil) held in front of the eyepieces at 30
cm must be clearly focused.
5. Using the Delivery Mirror Control Knob, adjust the optics so that the light spot
is centered vertically in your fi eld of view, then tighten the Observation Optics
Adjustment Knob.
6. If the light spot is not centered horizontally, adjust the headband left or right
accordingly.
7. Adjust the pupil distance setting by viewing the light spot alternately with the left eye
then the right eye, and sliding the eyepieces so that the spot is centered within your
eld of view.