User's Manual Part 5

6.8
8065751131
Operation
Operation of the Doctor Protection Filter
Before operating the slit lamp, the
Doctor Protection Filter cable must be plugged in and
rmly attached to the rear panel of the
PurePoint
laser system.
laser system.
The Doctor Protection Filter is operated by moving the lever from the unprotected
position to the fi lter protected (engaged) position. The 532nm laser treatment beam is not
operational until the fi lter lever is in the engaged position.
When using beam splitter accessories, the binoculars must fi rst be attached to the
beam
splitter (the beam splitter accessories are attached to the beam splitter on the protected
side of the Doctor Protection Filter assembly); the beam splitter is then attached to the
permanently installed Doctor Protection Filter (see Figure 6-9).
Positioning and
Focusing the Laser Beam
1. Move the joystick left and right to horizontally position the laser spot and illumination
slits.
2. Rotate the joystick to vertically position the laser spot.
3. Move the joystick forward and backward to focus the laser spot.
Adjusting the
Laser Position and Spot Size
1.
Following customary methods, position patient and place the contact lens on patient's eye.
2. Using the joystick control on instrument base, bring the selected area of treatment into
position/focus. If desired, lock instrument base in position with instrument base slide
set knob.
3. To position the laser spot, choose one or a combination of the following methods:
3.1 Using the joystick control, position the laser spot on the selected treatment area.
3.2 Using the micromanipulator, position the laser spot around the microscope center
eld of view.
3.3 Tilt the contact lens.
4. Use the laser spot size adjustment lever to set the laser spot size.
Laser Treatment of the Eye
1. The Doctor Protection Filter must be connected and in working order. To protect the
user's eyes, the fi lter must be in the engaged position prior to fi ring the treatment laser.
2. For laser spot sizes ranging from 50 µm to 500 µm, the focus is parfocal; i.e., the focus
of the laser spot lies in the focal plane of the microscope (see Figure 6-11). For laser
spots greater than 500 µm, the laser spot sizes are set by defocusing the laser; i.e., the
laser focus will not lie in the focal plane of the microscope).