User's Manual Part 4

4.2
8065751131
Inspect fi bers to ensure that they have not been compromised, i.e., chips, cracks, or loose
connectors.
The condition of the following system hardware components must be checked
periodically to identify any fault that may affect system operation:
Chassis appearance.
Operation of controls and indicators.
State of the fi bers and connecting cables.
Damaged hardware must be replaced to ensure safe operation. Call Alcon Technical
Services for assistance.
Mirror and Lens Cleaning
The mirrors and lenses of the LIO headpiece and Slit Lamp adaptation must be
kept clean and unscratched. Cleaning them requires special care and the following
materials:
Standard lens cleaning paper
Methanol of spectrographic quality.
The following tips will aid you in cleaning the optics:
Use each piece of cleaning paper only once.
Move the cleaning paper across the optic surface from one end to the other in one
continuous motion. Discard the cleaning paper and use a new piece for the next cleaning
pass.
Do not use a back and forth rubbing motion on the optic surface.
CAUTION
Care and cleaning operations must be performed with the instrument turned off and
power cord disconnected. Use only optical quality paper and spectroscopic quality
methanol when cleaning the mirrors and lenses, otherwise the optics could be
scratched and their coatings destroyed.
FUSE REPLACEMENT PROCEDURE
NOTE: Use only the recommended fuses for the
PurePoint
Laser as listed on
PurePoint Laser as listed on PurePoint
the fuse label.
1 Turn system power off and disconnect power cord
from the PurePoint ™ Laser before changing fuses.
2 Remove the fuse clip from the fuse holder using a
small screwdriver.
3 Inspect fuses in holder for damage or a burnt
connection.
4 Place new fuses in each side of holder in fuse clip
(replace with fuses rated T5A/250V or contact your
local Alcon representative).
5 Replace fuse clip into the fuse holder. Close fuse
holder.
Fuse