User's Manual Part 3
8065751131 3.9
Figure 3-5 Transition from Laser On to
Standby because of a Warning
4.5 Repeat the fi ring procedure as often as necessary, making adjustments to power
output and duration as appropriate to complete the treatment session.
If the system transitions back to Standby mode because of a warning condition, the
parameters section color is changed and a warning message is displayed as shown in
Figure 3-5. This mode change will occur if any of the following events occurs while
the laser is fi ring:
• Standby/Ready button pushed.
• Footswitch unplugged.
• Laser fi ber disconnected from
selected port.
• Remote Interlock disconnected
port.
• Dr. Filter connected, disconnected,
or disengaged.
• System fault.
To resume normal operation,
perform the action(s) displayed in
the warning message.
4.6 When the treatment is completed, release the footswitch and press the Standby/
Ready key. The Standby LED illuminates and the system is placed in standby
mode.
5 TURN OFF SEQUENCE
5.1 Turn the key to the OFF (O) position and, for safety reasons, remove the key.
NOTE: The emergency switch on the front panel should only be used in an
emergency. After using the emergency switch, pull it back to its initial
position to restore power and start the instrument.
5.2 Place the power switch on the rear of the system to the OFF (O) position.