User's Manual Part 3

3.6
8065751131
3.7 Set the desired inter-pulse time by turning the Inter-Pulse Time knob to the
desired setting in Repeat mode. To select Single Shot mode press the knob in or
turn the knob clockwise past the maximum 1 second inter-pulse time. In Single
Shot mode, the footswitch must be depressed for each shot. In Repeat mode, the
shots are fi red in a regular sequence while the footswitch is depressed and stops
when the footswitch is released.
3.8 Set the aiming beam intensity to the desired level by turning the Aiming Beam
Intensity knob.
WARNING!
Do not attempt treatment if aiming beam is not present. Patient injury may
occur.
NOTES: The footswitch must be released to proceed to Ready Mode. If the
footswitch is depressed during power-up or when switching from
Standby to Ready mode, the “Release Footswitch” message is
displayed.