User's Manual Part 3

3.4
8065751131
3 SYSTEM POWER UP AND SET UP
NOTE: Be sure to read all prompts on the display.
3.1 Turn the key to the ON (horizontal) position. Figure 3-2 shows the sequence of
screens displayed during initialization.
The front panel display illuminates the background LED’s and initializes the
LCD screen.
If no laser fi ber is connected to a port, a prompt to insert a probe is displayed
until one is connected.
During initialization of the system, the delta between the set point and actual
temperature of the laser engine is shown as it approaches nominal operating
temperature. The user can perform the following setup operations during
initialization: inserting probes, inserting LIO illumination power, set laser parameters
if an identifi ed probe is inserted, and view or change system settings.
NOTES: If the optical fi ber is not connected, the aiming beam will not turn on
and the system will continue to display: “Please Insert Probe .”
Display while initializing,
nothing plugged in the front
panel. No laser parameters
are set.
An identifi ed slit lamp fi ber has
been plugged-in and system
requests Dr. fi lter confi rmation.
Current footswitch
confi guration is displayed.
Fully initialized display
(Operating Display) with gray
background that indicates
Standby mode.
Figure 3-2 Display During Initialization
Display after warmup is
complete, system requests
probe to be inserted.