User's Manual Part 3

3.2
8065751131
WARNINGS!
The Slit Lamp must be equipped with a special Alcon Slit Lamp adaptation. This
adaptation is available for many of the existing Slit Lamps. (Reference section
six for list of adaptations to be used with existing slit lamps.) If peripherals are
not correctly connected and confirmed by the operator, the operator and patient
will be exposed to hazardous radiation.
Refer to Section Six for instructions on installing the Doctor Protection Filter.
The operator may have a colored view through the Doctor Protection Filter due
to blocking of the green 532 nm wavelength light. Newer filters will have less
color blocking. It is the operators responsibility to properly install the Doctor
Protection Filter. Alcon shall not be held liable for problems caused by improper
installation of the Doctor Protection Filter.
If a tethered manual Doctor Protection Filter is in the “not engaged” position the
prompt on the
PurePoint™
LCD display must read “Please engage Dr. Filter.” If
not, the operator must discontinue using the system and notify Alcon Technical
Services for assistance.
When using beam splitter accessories, the ocular stereo microscope head must first
be attached to the beam splitter (the beam splitter accessories are attached to the
beam splitter on the protected side of the Doctor Protection Filter assembly); the
beam splitter is then attached to the permanently installed Doctor Protection Filter.
Improper installation could cause injury to the operator and/or the patient.
It is the user's responsibility to ensure that non-RFID probes are correctly
identified.
Do not use assessories with a fiber or connector that have been compromised.
2.7
SLIT LAMP CONNECTION:
Connect the fi ber optic connection from the slit
lamp zoom to a laser fi ber port on the front panel (see Figure 3-1)
NOTE: When removing the fi ber of a slit lamp terminal or an LIO terminal, be
sure to secure the dust cover on the front panel fi ber port.
2.8
LIO CONNECTION:
2.8.1 Connect LIO fi ber to a laser fi ber port on the
PurePoint™
front panel.
2.8.2 Insert LIO power cord into LIO illuminator power port on front panel.
WARNING!
Endoprobes are for single-use only. Microbial or prion infection may occur if
re-used.
2.9
ENDOPROBE CONNECTION:
Connect fi ber to a laser fi ber port on the
PurePoint™
front panel.
NOTE: Refer to section six for additional information on setting up and using
these accessories.
2.10 Ensure Red Emergency Switch is pulled out. Press only in case of emergency.