User's Manual Part 2
8065751131 2.7
REAR PANEL DESCRIPTION
Figure 2-4 The
PurePoint™
Laser Rear Panel
Doctor Protection Filter Ports
Two ports are provided to connect Doctor's fi lters. If more than one fi lter is needed,
both ports can be used at the same time. There is no direct correspondence between
the laser ports and Dr. Filter ports.
If the Doctor Protection Filter is not engaged, the warning message “Please Engage
Dr. Filter” is displayed on the LCD. This message is displayed every time a port
with a slit lamp or endoprobe fi ber connected is selected. Acknowledging the prompt
removes the warning message and allows normal operation to continue.
WARNINGS!
The operator may have a colored view through the Doctor Protection Filter due
to blocking of the 532 nm wavelength (green). Newer filters are more transparent
and provide a less-discolored view.
The operator must be careful to avoid potential secondary reflections. Therefore,
the treatment room should be approved by a qualified laser safety officer.
Footswitch Connection
Either a standard footswitch or a multi-function footswitch is connected to the system
through this port. The multi-function footswitch provides additional controls to
change laser power and switch between Standby and Ready modes.
Power Cord Connector and Main Power Switch
The power cable for the system is connected in this module and the power switch is
located above the power connection. If the Emergency Switch is pressed, the power to
the system is shut off even if the Main Power Switch is ON.
Dr. Filter Ports
Remote Interlock
Remote Interlock
Connector
Laser Status
Connector
Connector
Footswitch Connector
Tether Port
Main Power Switch
Power Cord Connector