User's Manual Part 2

8065751131 2.3
A message is displayed as shown in Figure 2-3 prompting
the user to identify the probe as one of three instruments
displayed in the menu options.
After the user has identifi ed the probe:
The Port Illumination Ring is illuminated in the color
green if the port is the only port with an identifi ed probe.
The Port Illumination Ring is not lighted if the other port
already contains an identifi ed probe. The port with the
newly-identifi ed probe is not the selected port. The Port
Selection Button must be pushed for the probe to be
selected.
Disabled Port Indication Color:
RED
If the
PurePoint™
software determines a port cannot be
used due to a mechanical or software fault, the Port Illumination Ring is illuminated
in the color red.
Port Selection Buttons
When each port has an identifi ed probe, pressing one of the Port Selection Buttons
selects the associated port. The Port Selection Buttons are active only when there are
identifi ed probes in each port. In all other cases (such as an unidentifi ed probe in one
port, only one port being used, no identifi ed probes in a port, etc.) the
PurePoint™
software will take the correct action.
When a selection is made, the LED surrounding the desired port illuminates green. If
selecting a port for the fi rst time after starting the system with an identifi ed endoprobe
or slit lamp attachment, a screen is displayed asking the user to confi rm that the Dr.
Filter is in place.
Emergency Switch
Pressing the Emergency Switch turns off the laser system immediately. To restore
power, pull the Emergency Switch to the out position. The system will return to the
Standby mode.
Keyswitch
The Keyswitch is a two position switch where the key can be removed only in OFF
position. In the OFF position, the power to the laser is shut off. Switching to the ON
position turns the laser system on and the system will display the Standby screen.
Figure 2-3
Unidentifi able Probe
Selection Screen