User's Manual Part 1

8065751131
vii
IMPORTANT NOTICE
Equipment improvement is an on-going process and, as such, changes may be made to the
equipment after this manual is printed.
Pay close attention to
WARNINGS
and
CAUTIONS
in this manual.
WARNINGS
are written
to protect individuals from bodily harm.
CAUTIONS
are written to protect the instrument
from damage. Illustrations contained in this manual are for reference only.
It is recommended that maintenance be performed by a qualifi ed Alcon Field Engineer.
Alcon Surgical shall not be liable for any damage resulting from failure to comply with the
enclosed instructions.
Alcon reserves the right to change specifi cations without further notice.
CAUTION
U.S. Federal Law restricts this device to sale by or on the order of a physician only.
WARNINGS!
Use of controls or adjustments, or performance of procedures other than those specifi ed
herein may result in hazardous laser radiation exposure.
A qualifi ed technician must perform a visual inspection of the following components
every twelve months. In case of a defi ciency, do not use the system; call Alcon Technical
Services.
- Warning Labels
- Power Cord
- Fuses
A qualifi ed technician must check ground continuity and both polarities for leakage
current every twelve months to ensure they are within the applicable standard (for
example: EN 60601-1/IEC 601-1). Values must be recorded, and if they are above the
applicable standard, or 50% above your fi rst measurement, do not use the system; call
Alcon Technical Services.
WARNING!
Use of accessories and cables other than those provided may result in increased
emissions or decreased immunity of the system. Portable and mobile RF
communications equipment can affect this medical electrical equipment.
Comments or corrections concerning this manual should be addressed to:
Alcon
Technical Services Group
PO BOX 19587
Irvine, CA, USA 92623
All rights reserved. No part of this manual may be reproduced, transmitted, or stored in
a retrieval system, in any form or by any means; photocopying, electronic, mechanical,
recording, or otherwise; without prior written permission from Alcon Laboratories, Inc.