User's Manual Part 1

1.12
8065751131
Utility Requirements
Electrical requirement: The
PurePoint™
Laser has a power supply that operates at
100-120 V and 220-240 V input ranges at 50/60 Hz. A properly grounded, standard
plug is the only requirement.
Electrical Connections
CAUTION
Before turning the instrument ON for the fi rst time after receipt of the system, wait one
hour for the components and optics to normalize to avoid possible condensation that
may have occurred during shipping.
Use only <HAR> power cord with a minimum of 10 Amp rating.
Before connecting the main plug verify that:
The Main Switch on the back panel is in the OFF (O) position.
The key is in the off (vertical) position, or has been removed.
The Remote Plug or the interlock cable is connected on the rear panel.
Optical Connections
Optical connections vary in relation to the procedure to be performed. Different
peripherals can be connected to the output ports. These peripherals are:
Slit Lamp adaptation
Laser Indirect Ophthalmoscope (LIO)
Endoprobe/Aspirating Endoprobe/Illuminated Endoprobe
The procedures for connecting these peripherals are contained in Section Three:
Operating Instructions.
Figure 1-5 Remote Connector/Door Lamp Circuit Diagram
REMOTE
CONNECT
9-PIN D-CONNECTOR
CUSTOMER SUPPLIED
DOOR SWITCH (OPTIONAL)
TO DRIVE DOOR LAMP (OPTIONAL)
1
9
3
NONC
COM
1
6