User's Manual

8065751772 1.5
Table 1-1
Guidance and Manufacturer's Declaration - Electromagnetic Emissions - The Centurion
®
Vision System is intended for use in the electromagnetic environment specied below. The
customer or the user of the Centurion
®
Vision System should assure that it is used in such
an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage uctuations/
Flicker emissions
IEC 61000-3-3
Electromagnetic Environment-Guidance
The Centurion
®
Vision System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The Centurion
®
Vision System is suitable for use in all establishments
other than domestic and those directly connected to a low voltage power
supply network that supplies buildings used for domestic purposes.
The EMC Statement provides guidance on steps to take in case of
electromagnetic interference.
Compliance
Group 1
Class A
Class A
Complies
Users should be aware of known RF sources, such as radio or TV stations and hand-held
or mobile two-way radios, and consider them when installing a medical device or system.
Portable and mobile RF communications equipment such as cellular telephones can affect
medical electrical equipment (see Table 1-3 for recommended separation distances).
Be aware that adding accessories or components, or modifying the medical device or
system, may degrade the EMI performance. Consult with qualied personnel regarding
changes to the system conguration.
WARNINGS!
The use of accessories, transducers, and cables other than those specied, with
the exception of transducers and cables sold by the manufacturer of the system as
replacement parts for internal components, may result in increased emissions or
decreased immunity of the system.
The system should not be used adjacent to, or stacked with, other equipment; and that if
adjacent to or stacked use is necessary, the system should be observed to verify normal
operation in the conguration in which it will be used.
MAGNETIC AND ELECTRICAL INTERFERENCE - Magnetic and electrical elds are
capable of interfering with the proper performance of the device. For this reason make
sure that all external devices operated in the vicinity of the device comply with the
relevant EMC requirements. X-ray equipment, magnetic resonance tomography (MRT),
nuclear magnetic resonance (NMR), or magnetic resonance imaging (MRI) devices are
possible sources of interference as they may emit higher levels of electromagnetic
radiation. See the Magnetic Resonance Unsafe icon in Figure 1-2.