User's Manual

8065751772 1.21
Consumable Packs
Consumable items used with the Centurion
®
Vision System during surgery are
designed to be used once and then discarded, unless labeled otherwise.
All Centurion
®
packs contain Directions for Use (DFU). It is important to read and
understand the DFU’s prior to use.
In all cases, the instrument setup instructions contained in the manual should be
thoroughly understood prior to using any of the pack congurations.
NOTE: If an inconsistency exists between the instructions in the operator’s
manual and the Directions For Use (DFU) supplied with a consumable pack or
accessory, follow the DFU.
WARNINGS!
Mismatch of consumable components and use of settings not specially adjusted for
a particular combination of consumable components may create a patient hazard.
Do not use packs that have exceeded the expiration date.
Sterile disposable medical devices should not be reused! These components have
been designed for one time use only; do not reuse.
Potential risk from reuse or reprocessing the following products labeled for single
use include:
Bipolar Coagulation Instruments - Thermal injury or electrical shock caused by a
damaged bipolar instrument, and foreign particle introduction into the eye.
Fluid Management Components - Fluid path leaks or obstruction resulting in reduced
uidics performance, and foreign particle introduction into the eye.
Phacoemulsication Tips - Reduced tip cutting performance, presence of tip burrs,
uid path obstruction, and foreign particle introduction into the eye.
Vitreous Cutting Instruments - Reduced vitreous cutting performance, uid path
obstruction, and foreign particle introduction into the eye.
The equipment used in conjunction with the Alcon disposables constitutes a complete
surgical system. Use of disposables other than Alcon disposables may affect system
performance and create potential hazards, and if it is determined to have contributed
to the malfunction of the equipment under contract, could result in the voidance of
the contract and/or invoicing at prevailing hourly rates.
Perform visual inspection of accessories for burs or bent tips prior to use.
Do not remove the FMS during the procedure.
Do not use any Vitrectomy probes that have not been approved for use on the
Centurion
®
Vision System.