User's Manual Part 4
2.84 8065751025
Constellation
®
Vision System
The following controls are available only in the Advanced view:
Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is
reached, the vacuum level can no longer be increased. If the Flow Limit is turned
off, the vacuum level is limited only by the Start and End values.
Flow Limit Up/Down Buttons - Sets the maximum flow rate.
Flow Limit Progress Bar and Numerical Value - Displays the current flow rate.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Flow Limit Configuration Button - Displays the configuration popup.
Flow Limit Control Drop Down List - Determines whether the flow limit is fixed
or linear. In Fixed, the limit is constant regardless of the flow level. In Linear, the
limit varies depending on the flow level.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Aspiration Panel:
Flow End Up/Down Buttons - Sets the flow level at full treadle.
Flow Progress Bar and Numerical Value - Displays the current flow level.
Aspiration Configuration Button - Displays the configuration popup.
Vacuum Limit On/Off Button - Turns vacuum limit on or off. When the maximum
•
•
•
•
•
•
•
•
•
•
•