User's Manual Part 4

2.92 8065751025
Constellation
®
Vision System
set the pressure at Start and Close.
- When Start is selected, the Start pressure is applied to the port and the Start Up/
Down buttons are active on the Pressure control to allow adjustment of the Start
pressure.
- When Close is selected, the Close pressure is applied to the port and the Close
Up/Down buttons are active on the Pressure control to allow adjustment of the
Close pressure.
- Pressing the "Save Calibration Values" button saves these values.
Forceps Panel:
Bimanual On/Off Button - Turning Bimanual mode on allows the control of the
pneumatic forceps while the system is in Scissors mode. All controls except for
the footswitch function the same as when the system is in Forceps mode (see
Forceps mode for a description).
Viscous Fluid Control (VFC) Mode
The VFC (Viscous Fluid Control) mode provides pressure at the front panel VFC
connector for fluid injection (i.e., silicone oil), or vacuum for extraction, through the
VFC tubing set to a syringe. Using vacuum, the VFC mode also provides a means of
extruding fluid through the VFC consumables.
CAUTION
Always use Alcon-supplied Viscous Fluid Control Kits and follow all Directions
for Use. Do not use VFC without the plunger/stopper supplied with the kit. Do
not aspirate fluids directly into the console; this will cause damage to the console,
increase risk of electrical shock, and void all warranties.
Viscous Fluid Control (VFC) Mode: Extract
Submode
Figure 2-69 Surgery Screen: VFC Mode - Extract Submode