User's Manual Part 4

2.88 8065751025
Constellation
®
Vision System
In “Standard” display mode, only the Vacuum mode controls are available. The
controls for each panel in Laser mode have the following functions:
Vacuum Panel:
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. The Start
value is always zero.
Vacuum Progress Bar and Numerical Value - Displays the current vacuum level
Vacuum Configuration Button - Displays the configuration popup. Simplifies
large value changes by providing slider controls. Also provides panel specific
configuration capabilities such as Drainbag Change.
The following controls are available only in the Advanced view:
Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is
reached, the vacuum level can no longer be increased. If the Flow Limit is turned
off, the vacuum level is limited only by the Start and End values.
Flow Limit Up/Down Buttons - Sets the maximum flow rate.
Flow Limit Progress Bar and Numerical Value - Displays the current flow rate.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Flow Limit Configuration Button - Displays the configuration popup.
Flow Limit Control Drop Down List - Determines whether the flow limit is fixed
or linear. In Fixed, the limit is constant regardless of the flow level. In Linear, the
limit varies depending on the flow level.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Aspiration Panel:
Flow End Up/Down Buttons - Sets the flow level at full treadle.
Flow Progress Bar and Numerical Value - Displays the current flow level.
Flow Configuration Button - Displays the configuration popup.
The following controls are available only in the Advanced view:
Vacuum Limit On/Off Button - Turns vacuum limit on or off. When the maximum
rate is reached, the flow can no longer be increased. If the Vacuum Limit is turned
off, the flow is limited only by the Start and End values.
Vacuum Limit Up/Down Buttons - Sets the maximum vacuum value.
Vacuum Limit Progress Bar and Numerical Value - Displays the current vacuum
level.
Vacuum Limit Configuration Button - Displays the configuration popup.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Laser Panel:
Laser Mode - Selects laser Standby or Ready mode.
Port Selection - Selects the port used for treatment.
Treatment Mode - Selects the Treatment mode: Repeat, Single Shot, or
Continuous
Power Up/Down Buttons - Adjusts the laser power level.