User's Manual Part 4
8065751025 2.85
Constellation
®
Vision System
rate is reached, the flow can no longer be increased. If the Vacuum Limit is turned
off, the flow is limited only by the Start and End values.
The following controls are available only in the Advanced view:
Vacuum Limit Up/Down Buttons - Sets the maximum vacuum value.
•
•
Figure 2-64 Surgery Screen: Extrusion Mode
Vacuum Limit Progress Bar and Numerical Value - Displays the current vacuum
level.
Vacuum Limit Configuration Button - Displays the configuration popup.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Extrusion Mode
In Extrusion Mode the Surgery screen contains the Vacuum panel necessary for
controlling an extrusion handpiece. The vacuum pressure is always controlled linearly
and vacuum cannot be turned off.
In “Standard” display mode, only the Vacuum mode controls are available. The
controls for each panel in Extrusion mode have the following functions:
Vacuum Panel:
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. The Start
value is always zero.
Vacuum Progress Bar and Numerical Value - Displays the current vacuum level
Vacuum Configuration Button - Displays the configuration popup. Simplifies
large value changes by providing slider controls. Also provides panel specific
configuration capabilities such as Drainbag Change.
The following controls are available only in the Advanced view:
Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is
reached, the vacuum level can no longer be increased. If the Flow Limit is turned
•
•
•
•
•
•
•