Same as Burst submode.
Ultrasound Panel:
Phaco Power End Up/Down Buttons - Sets the power at full treadle.
Phaco Power Progress Bar and Numerical Value - Shows the current power value.
Phaco Power Limit Control Drop Down List - Determines whether the power
is fixed or linear. In Fixed the power is constant regardless. In Linear the power
varies depending on the treadle position.
Smart Pulse - Indicator for smart pulse becomes visible when power is greater
than 0 and pulse on time is less than 20ms
Phaco Power Configuration Button - Displays the configuration popup. Simplifies
large value changes by providing slider controls.
Phaco Pulse On Time Up/Down Buttons - Sets the Phaco pulse on time.
Phaco Pulse On Time Progress Bar and Numeric Value - Shows the current Phaco
pulse on time.
Phaco Pulse On Time Limit Control Drop Down List - Determines whether the
pulse on time pulse type limit is decreasing, fixed, or increasing. In Fixed the limit