User's Manual Part 3

8065751025 2.73
Constellation
®
Vision System
initiate a drainbag change.
Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is
reached, the vacuum level can no longer be increased. If the Flow Limit is turned
off, the vacuum level is limited only by the Start and End values.
Flow Limit Up/Down Buttons - Sets the maximum flow rate.
Flow Limit Progress Bar and Numerical Value - Shows the current flow level.
Flow Limit Configuration Button - Displays the configuration popup.
Flow Limit Control Type Drop List - Determines whether the flow limit is fixed
or linear. In Fixed the limit is constant regardless of the flow level. In Linear the
limit varies depending on the flow level.
Flow/Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Ultrasound Panel:
Power End Up/Down Buttons - Sets the power at full treadle.
Power Progress Bar and Numerical Value - Displays the current power value.