The controls for each panel in Burst submode have the following functions:
Aspiration Panel:
Flow End Up/Down Buttons - Sets the flow level at full treadle.
Dynamic Rise Drop List - Select for the rise value. The rise value determines
the slope of the increase applied.
Flow Configuration Button - Displays the configuration popup.Simplifies large
value changes by providing slider controls. Also provides the capability to initiate
a drainbag change.
Vacuum Limit On/Off Button - Turns vacuum limit on or off. When the maximum
rate is reached, the flow can no longer be increased. If the Vacuum Limit is turned
off, the flow is limited only by the Start and End values.
Vacuum Limit Up/Down Buttons - Sets the maximum vacuum value.
Vacuum Limit Control Type Drop List - Determines whether the vacuum limit is
fixed or linear. In Fixed the limit is constant regardless of the vacuum level. In
Linear the limit varies depending on the vacuum level.
Vacuum Limit Configuration Button - Displays the configuration popup.
Flow/Vacuum Toggle (Advanced view only): Switch between Flow Mode and
Vacuum Mode.
Vacuum Panel:
The following controls are available only in the Advanced view only by switching
the Function Mode from Flow to Vacuum:
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle.
Vacuum Configuration Button - Displays the configuration popup. Simplifies
large value changes by providing slider controls. Also provides the capability to
•
•
•
•
•
•
•
•
•
•