User's Manual Part 2
2.44 8065751025
Constellation
®
Vision System
Basic Setup: The Handpiece Setup Panel
This panel shows setup information related to handpiece selections. Figure 2-33
shows the Handpiece Setup Panel and the Detailed Handpiece Setup Popup screen.
Figure 2-33 The Handpiece Setup Panel
Type Button - Pressing the Type button on the panel displays the list of handpieces
that may be used with the system. When the system detects that the user has plugged
in the handpiece, the handpiece name is automatically selected and the “connected”
icon appears next to the selection list.
The following handpiece types are selectable from the drop list:
Extrusion
Extrusion SoftTip
Fragmentation*
Irrigation/Aspiration
Phaco*
None
Handpieces that are marked with an asterisk (*) are automatically detected by the
system. When the system detects connection of a handpiece, the user will not be able
to select another handpiece from the touchscreen (the user list is locked).
Tip Button - Pressing the Tip button displays the tips that the operator can choose
for the currently selected/inserted handpiece. The tip selection button is hidden if the
currently selected handpiece is “None”
Expand Button - Pressing the expand button opens the detailed setup popup that
is covered later in this section.
Handpiece status is displayed at the bottom of the setup panel and indicates whether
the probe has been primed and/or tuned. A "√"indicates that priming and/or tuning
has been completed, while an "X " indicates the it has not.
•
•
•
•
•
•