User's Manual Part 2
8065751025 2.27
Constellation
®
Vision System
Menu Bar: Options - DOCTOR SETTINGS - LASER
The settings available in the Laser settings screen are for setting the Revert to
Standby Timeout and Footswitch configuration (see Figure 2-15).
Revert to Standby Timeout
This setting determines the time period, in minutes, after which the system will
revert to the Standby mode of operation due to inactivity. Inactivity is defined as no
footswitch activation during the time period. The setting can be adjusted to 2, 5 or 10
minutes.
Footswitch Configuration
The Side Switch settings allow the user to enable the side switches to control laser
power and transition the system between Ready and Standby.
Power Control Only
This selection allows the side switches to increment (left switch) or decrement (right
switch) the treatment laser power.
Power & Ready <-> Standby
When this selection is enabled, the user can transition from Standby to Ready by
pressing the right side switch longer than 1.5 seconds until a tone sounds. Similarly,
pressing the left side switch longer than 1.5 seconds (until a tone sounds) will
transition the system from Ready back to Standby. Pressing the side switches for
intervals shorter than 1.5 seconds will increase or decrease treatment laser power.
Standard
This setting disables the side switches entirely so the footswitch can only be used for
firing the laser. Power control and switching from Standby to Ready mode can only
be done from the front panel of the system.
Figure 2-15 Doctor Settings - Laser Screen