User's Manual Part 2
8065751025 2.25
Constellation
®
Vision System
Figure 2-13 Doctor Settings - Footswitch Treadle Screen
Menu Bar: Options - DOCTOR SETTINGS - FOOTSWITCH TREADLE
The Footswitch Treadle screen shown in Figure 2-13 enables the user to program the
treadle vibration, detent firmness, and spans. Depending on the current surgical mode
and sub-mode, the treadle has either 2 or 3 positions and the user can program the
treadle travel of each position.
2-Position Footswitch Transition Position - Sets the transition position (in % of
treadle travel). Valid range: 20% - 80%
3-Position Footswitch Transitions
- Point 1 - Sets the transition position (in % of treadle travel) for transition
point 1. Valid range: 15% - 31%
- Point 2 - Sets the transition position (in % of treadle travel) for transition
point 2. Valid range: 50% - 100%
Vibration Enable/Disabled - Allows the user to enable/disable treadle vibration
(for both treadle up and treadle down).
Vibration Frequency - Adjusts the vibration frequency of the treadle when it goes
through detents.
Treadle Firmness - Adjust the firmness of the treadle in a range from 0 to 100%.
•
•
•
•
•