Product Specifications
General Information
Product Use: Indoor use only
Operating Temperature Range: 32-122F (0-50C)
Recommended Storage Temperature
Range:
32-122F (0-50C)
Relative Humidity Range: 0%-80%
Polluction Degree: Degree 2; Normal
Altitude: For use up to 14,000 ft.
Measurement Range: 0.000 – 0.400 BrAC
Breath Sample Volume: 1.5L
Accuracy: +/-0.005 up to 0.100 BrAC
+/-5% from 0.100 to 0.400 BrAC
Alcohol Sensor: Proven Fuel Cell Technology
Automated Facial Intelligence (AFI): Proprietary 1-to-1 facial template
matching
Battery: Rechargeable Lithium Ion
Input Power: 5VDC @ 2.0A, 10.0 W (0.4A max)
Charger: 5.0V / 2.0A USB-C Plug
Best results are obtained with no more than +/- 15 degrees head roll, pitch, or
yaw; neutral face expression; no sunglasses or heavy-framed glasses; and no
hair in the face.
Maintenance, Inspection, and Device
Return Conditions
The device conducts a series of diagnostic checks each time it wakes up.
If the device fails a diagnostic check, the device should be removed from
service and returned to the manufacturer for maintenance.
The device should be removed from service and returned to the
manufacturer if the device fails to calibrate to the established
specications.
The device should be removed from service and returned to the
manufacturer if requested by an authorized AMS representative.
A physical inspection of the device should be conducted before testing by
the device operator to ensure there are no signs of damage.
Repairs to the device should only be conducted by trained AMS
technitions.