Calibration/Calibration Check
Issue Indication Recommended Action
Calibration Failed The calibration or
calibration check
does NOT meet
specications.
Follow recommendations
displayed on device:
• Check rubber tubing
connections
• Ensure breath tube
connection to the device is
secure
• Check the canister
pressure
Cal-Check Failed The calibration check
does NOT meet
specications.
Follow recommendations
displayed on device:
• Check rubber tubing
connections
• Ensure the breath tube
connection to the device is
secure
• Check the canister
pressure
Device Timed Out The device did NOT
detect gas for 15
minutes after being
turned on.
Re-initiate the Calibration/
Calibration Check in SCRAMNET
and begin the ow of gas from
the calibration canister within 15
minutes of initiation.
Device message
still reads "Start
Flow of Gas" even
when the regulator
button is held
down.
The device is not
detecting the ow
of gas.
Ensure there is adequate
gas pressure to complete a
calibration or cal-check.
Ensure the correct regulator
is being used. The RB Pro
regulator delivers 4L of as per
minute.